ENProcessing Instructions.Cleaning – so easy.
General instructions.6. Starting the cleaning programme of the Dental Washer Disinfector according to manufacturer´s instructions.7. Removing ti
Maintenance.Reassembly of instruments (p. 14 section Reusa-bility of surgical instruments).If possible, instrument oils should not be used. If oil is
General instructions. Proof of basic suitability for effective steam sterilization was provided by an independent, accredited test laboratory using a
Correct storage.After sterilization the tioLogic© easyClean has to be stored dry and dust free in the sterilization container.Material resistance.When
Torque ratchet.Disassembly. Fully unscrew the adjusting screw from the torque ratchet handle and remove the springs. Leave the stopper on the spring.L
Insert the spring with the stopper towards the front into the ratchet handle and tighten the adjusting screw slightly.Check the function after assemb
989-998-20 Printed by Dentaurum Germany 03/13 For more information on our products and services, please visit www.dentaurum.deDate of infor
Another benefit of tioLogic© easyClean: during implantation, the instruments can be put back in the corresponding slots after use. Hence, each instrume
Content.tioLogic© easyClean Wash-TrayCleaning – so easy 6General instructionsApplicationCleaning and desinfectionPreconditioningCare, checking, mainte
5
6The tested tioLogic© easyClean can be excellently cleaned without mechanical pre-cleaning.1Developed in collaboration with:1 Cleaning investigation
7The tioLogic© easyClean is a Wash-Tray that contains all the rotary instruments and accessory components needed for implantation. They are arranged
tioLogic© easyClean instruments can be put back in the tray when they are dirty. Prior to cleaning in the Dental Washer Disinfector, all instruments t
Mechanical cleaning / disinfection (disinfector or cleaner / disinfector).When choosing a disinfector, ensure that the efficacy of the disinfec